Non-ferrous Water Cooling System

GRE were chosen to design and manufacture the secondary cooling element of an MRI-guided radiation therapy system for a world leading manufacturer of ground breaking oncology equipment.

 

This machine was first-in-class and the cooling system needed to be completely bespoke, not only because of the magnetic and radioactive nature of the equipment, but also due to the spherical shape of the scanner. With this in mind, there were very limited material selection options. Plastics were used wherever possible, but again had to be carefully considered due to the high radiation they would be subjected to.

Key Features:

  • 14kW design cooling capacity
  • All materials non ferrous (i.e. non magnetic)
  • Entire system subjected to high levels of radiation
  • Designed as a curved section to form part of a segment within the entire machine
  • Customised control package
  • Wifi network access (due to constant revolution of the unit)
  • At certain times during the rotation, air to the heat exchanger could be restricted and therefore a backup cooling store was also designed in using thermal wax to store latent cooling effect (system recognises this, automatically adapts to use or rebuild the stored cooling)

 

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

You can adjust all of your cookie settings by navigating the tabs on the left hand side.

For more information on our cookies, visit the link at the bottom of our website.